FCE 99M-PVS-R

CALIBRATED LEAK EN 14624

CALIBRATED LEAK

R134a, R1234ze, R1234yf, R600a, R290

The FCE 99M-PVS-R calibrated leak is intended for sensitivity checking and calibration of field and production leak detectors operating by sniffing at atmospheric pressure, in accordance with the EN 14624 standard.

In the same range:

For an adjustable rather than fixed value, the FCE 24-PVA-DR has a built-in dual reservoir (expansion chamber and liquid-phase gas reservoir); it is available for HFC, HFO and HC refrigerant gases.

Compatible with:

The FCE 99M-PVS-R calibrated leak compliant with the EN 14624 standard enables sensitivity testing of refrigerant gas detectors: DXR 240, HRD 92 SC, HRD 92 E and MTD 920.

  • Standard leak rate of 3 g/yr or 5 g/yr at the gas vapour pressure at 20 °C.
  • Designed to be supplied with gas in the saturated vapour phase.
  • It has its own liquid-phase gas reservoir. SCHRADER-type ¼ SAE male filling valve.
  • Delivered with a calibration certificate and a pressure-temperature/leak rate curve.

Available for:

  • R134a, R1234ze, R1234yf, R600a, R290.
UseThe FCE 99M-PVS-R is a continuous-flow leak with its own liquid-phase gas reservoir. SCHRADER-type ¼ SAE male filling valve.
Leak rate

3 g/yr [±20%] at the vapour pressure of the gas at 20 °C.

5 g/yr [±20%] at the vapour pressure of the gas at 20 °C.

Other values on request.

Gas

R134a, R1234ze, R1234yf, R600a, R290.

Other gases on request.

Delivered withLeak delivered with its calibration certificate valid for 1 year and a pressure-temperature/leak rate curve.

Usage tips

  • Leak rate and temperature: the leak rate is given at the vapour pressure of the gas at 20 °C. At another temperature, the estimated value is shown on the pressure-temperature/leak rate curve supplied with the leak.
  • Liquid level: the vapour pressure, and therefore the leak rate, only remains stable as long as the reservoir contains liquid. The reservoir must therefore be refilled before it runs empty.
  • One gas per leak: the leak is calibrated for the gas stated on its certificate. Refilling with another refrigerant distorts the calibration and can destroy the leak.