FCE 99M-PVS-R
CALIBRATED LEAK EN 14624
CALIBRATED LEAK
R134a, R1234ze, R1234yf, R600a, R290
The FCE 99M-PVS-R calibrated leak is intended for sensitivity checking and calibration of field and production leak detectors operating by sniffing at atmospheric pressure, in accordance with the EN 14624 standard.In the same range:
For an adjustable rather than fixed value, the FCE 24-PVA-DR has a built-in dual reservoir (expansion chamber and liquid-phase gas reservoir); it is available for HFC, HFO and HC refrigerant gases.
Compatible with:
The FCE 99M-PVS-R calibrated leak compliant with the EN 14624 standard enables sensitivity testing of refrigerant gas detectors: DXR 240, HRD 92 SC, HRD 92 E and MTD 920.
- Standard leak rate of 3 g/yr or 5 g/yr at the gas vapour pressure at 20 °C.
- Designed to be supplied with gas in the saturated vapour phase.
- It has its own liquid-phase gas reservoir. SCHRADER-type ¼ SAE male filling valve.
- Delivered with a calibration certificate and a pressure-temperature/leak rate curve.
Available for:
- R134a, R1234ze, R1234yf, R600a, R290.
| Use | The FCE 99M-PVS-R is a continuous-flow leak with its own liquid-phase gas reservoir. SCHRADER-type ¼ SAE male filling valve. |
| Leak rate | 3 g/yr [±20%] at the vapour pressure of the gas at 20 °C. 5 g/yr [±20%] at the vapour pressure of the gas at 20 °C. Other values on request. |
| Gas | R134a, R1234ze, R1234yf, R600a, R290. Other gases on request. |
| Delivered with | Leak delivered with its calibration certificate valid for 1 year and a pressure-temperature/leak rate curve. |