FCH 99M-PVA-R150
ADJUSTABLE CALIBRATED LEAK WITH TANK
Helium, Argon, H2/N2, H2, R32, CO2, SF6
The FCH99M-PVA-R150 adjustable calibrated leak is intended for sensitivity control and calibration of production and laboratory leak detectors operating by sniffing at atmospheric pressure.In the same range:
For an adjustable leak supplied by an external bottle with pressure regulator (without built-in tank), the FCH 99M-PVA adjustable leak is the alternative for HFC HFO HC refrigerant gases, as well as CO2, H2/N2, Helium and Argon.
Compatible with:
The FCH 99M-PVA-R150 adjustable calibrated leak enables sensitivity testing of multi-gas production detectors MTD 920 and MTD 930, as well as the Helium and Argon leak detector TG EXPERT.
- Leak value adjustable in a value range depending on the applied pressure.
- Designed to be supplied with gas in the “dry” vapor phase.
- It has its own gas tank to fill in the “dry” vapor phase. 1/4FL filling needle valve.
- Delivered with a calibration certificate and a pressure/leakage flow correspondence curve.
Can be delivered for:
- Helium, Argon, H2/N2, H2, R32, CO2, SF6
Use | The FCH 99M-PVA-R150 leak is a continuous flow adjustable leak, it must be supplied with gas in the “dry” vapor phase. It has its own gas tank to fill in the “dry” vapor phase. 1/4FL filling needle valve. |
| Leak value | Leak value adjustable between 1 g/yr and 10 g/year approximately [+/-20%] depending on the pressure applied. The adjustment range depends on the version of the leak and the type of gas. Other values, contact us. |
| Gas | Helium, Argon, H2/N2, H2, R32, CO2, SF6. Other gases contact us. |
| Delivred with | Leak delivered with its calibration certificate valid for 1 year and a leak pressure/flow correspondence curve. |