FCE 24-PVA-DR

ADJUSTABLE CALIBRATED LEAK WITH RESERVOIR

R134a, R1234ze, R1234yf, R32, R290, R600a

The FCE 24-PVA-DR adjustable calibrated leak is intended for sensitivity checking and calibration of production and laboratory leak detectors operating by sniffing at atmospheric pressure, in accordance with the EN 14624 standard.

In the same range:

For a fixed value of 3 or 5 g/yr compliant with the EN 14624 standard, the FCE 99M-PVS-R EN 14624 calibrated leak, intended for certification, has its own reservoir; it is available for R134a, R1234ze, R1234yf, R600a and R290.

Compatible with:

The FCE 24-PVA-DR adjustable calibrated leak enables sensitivity testing of refrigerant gas detectors: DXR 240, HRD 92 SC, MTD 920 and MTD 930.

  • Leak rate adjustable over a range depending on the applied pressure.
  • Designed to be supplied with gas in the “dry” vapour phase.
  • It has a dual reservoir: an expansion chamber and a gas reservoir to be filled in the liquid phase. SCHRADER-type ¼ FL/SAE male filling valve.
  • Delivered with a calibration certificate and a pressure/leak rate curve.

Available for:

  • R134a, R1234ze, R1234yf, R32, R290, R600a
Use

The FCE 24-PVA-DR is an adjustable continuous-flow leak that must be supplied with gas in the “dry” vapour phase.

It has a dual reservoir: an expansion chamber and a gas reservoir to be filled in the liquid phase. SCHRADER-type ¼ FL/SAE male filling valve.

Leak rate

Leak rate adjustable between approx. 1 and 5 g/yr [±20%] depending on the applied pressure.

The adjustment range depends on the leak version and gas type.

Other values on request.

Gas

R134a, R1234ze, R1234yf, R32, R290, R600a.

Other gases on request.

Delivered withLeak delivered with its calibration certificate valid for 1 year and a pressure/leak rate curve.

Usage tips

  • Pressurisation: the leak, placed upright on a flat surface, is pressurised by opening the valve slowly, without exceeding the full scale of the pressure gauge.
  • Dry vapour supply: the expansion chamber of the dual reservoir delivers gas to the leak in the dry vapour phase. The leak rate then depends on the pressure read on the pressure gauge, not on temperature.
  • One gas per leak: the leak is calibrated for the gas stated on its certificate. Filling the reservoir with another refrigerant distorts the calibration and can destroy the leak.