HND 92 SC
TRACER GAS LEAK DETECTOR
FORMING GAS, N2/H2 TRACER GAS & NIDRON
The HND 92 SC is a portable gas leak detector for the detection of hydrogenated nitrogen gas (forming gas), nidron or N2/H2 tracer gas. Designed for leak testing by sniffing at atmospheric pressure.In the same range:
For portable use with graphic display and 2 detection modes (Measure and Track), the DXH 240 is the direct alternative on H2/N2 forming gas or Nidron tracer gas. For high-sensitivity tightness control on production lines or expert use, the MTD 920 provides automatic calibration on a calibrated leak and displays the estimated leak rate value.
Accessories:
For verification with a specific leak rate, use the FCH 99M-PVA-R150 adjustable calibrated leak with its built-in tank, delivered with a pressure/flow curve.
- Industrial detector for the detection of N2/H2 tracer gas.
- Graduated light bar allowing estimation of the detected leak.
Detection block :
- Selective detection block for N2/H2 tracer mixture.
Option : Lighting detection block.
| Functions | 5 sensitivity positions. Automatic zero on the ambient atmosphere. Contamination-saturation light. Interchangeable detection block. |
| Display | Graduated light bar with audible signal. |
| Sensor | Electrochemical. |
| Detected gases | Hydrogenated nitrogen, N2/H2 tracer gas mixture, NIDRON. |
| Sensitivity | 1.10-5 cc/s [equivalent approx. 1 g/yr refrigerant gases]. |
| Autonomy | 6h |
| Power supply | Built-in rechargeable batteries. Charging time around 12 hours. |
| Delivered with | Detector delivered complete in transport case with mains charger and instructions for use. |
| Option | Lighting block. |