MCB 97T3

HELIUM & ARGON CONCENTRATION METER

HELIUM & ARGON CATHAROMETER

The MCB 97T3 is a catharometer-type concentration meter for helium and argon. It measures the concentration of binary or pseudo-binary gas mixtures from 0.1% to 100%.

In the same range:

For portable, one-off helium or argon concentration measurement, the TG DISPLAY is the dedicated catharometer concentration meter, mainly used for enclosure (hood) leak testing.

  • Concentration reading from 0.1% to 100%.
  • ATEX non-saturating thermal conductivity measuring cell.
  • Rack 19″, 3U, 21TE
  • 3-digit digital display.

Calibration:

  • Pre-calibrated for helium.
  • Other calibrations available on request.
Functions

19″ rack-mount unit, 3U, 21TE.

2 cell mounting options (depending on version):

  • Remote cell connected by cable with gas circulation by natural convection.
  • Cell mounted on the rear of the housing, with cover and fittings for connecting gas supply tubes.
Display

3-digit digital display.

0–1 V analog output.

2 setpoint relays [+] and [-] with adjustable threshold, with NO/NC contact outputs.

CellATEX non-saturating thermal conductivity cell.
Detected gasesFactory-calibrated for helium; other gases on request (argon, etc.).
SensitivityConcentration reading from 0.1% to 100%.
Battery lifen/a
Power supplyMains supply 220 V, 50/60 Hz.
Delivered withSupplied complete with plug-in screw terminal blocks and user manual.
Optionn/a

Usage tips

  • Purging the sampling line: the reading is taken once the line has been fully purged. Any air ingress through a fitting dilutes the sample and distorts the measurement.
  • Sample quality: measurement accuracy depends on the precision of the calibration and of the zero. As the reading of a thermal conductivity cell can vary with the humidity of the mixture and the applied pressure, the sampled gas must be dry, at a stable flow rate and at atmospheric pressure (neither under vacuum nor overpressure).